Tabela PT100

Periodicamente, deve-se fazer uma verificação dos sensores de temperatura do equipamento, a fim de se evitar variações de leitura e erros de indicação.

 

T  (°C)R (ohm)T  (°C)R (ohm)T  (°C)R (ohm)T  (°C)R (ohm)T  (°C)R (ohm)
0100,0030111,6760123,2490134,70120146,06
1100,3931112,0661123,6291135,08121146,44
2100,7832112,4562124,0192135,46122146,81
3101,1733112,8363124,3993135,84123147,19
4101,5634113,2264124,7794136,22124147,57
5101,9535113,6165125,1695136,60125147,94
6102,3436113,9966125,5496136,98126148,32
7102,7337114,3867125,9297137,36127148,70
8103,1238114,7768126,3198137,74128149,07
9103,5139115,1569126,6999138,12129149,45
10103,9040115,5470127,07100138,50130149,82
11104,2941115,9371127,45101138,88131150,20
12104,6842116,3172127,84102139,26132150,57
13105,0743116,7073128,22103139,64133150,95
14105,4644117,0874128,60104140,02134151,33
15105,8545117,4775128,98105140,39135151,70
16106,2446117,8576129,37106140,77136152,08
17106,6347118,2477129,75107141,15137152,45
18107,0248118,6278130,13108141,53138152,83
19107,4049119,0179130,51109141,91139153,20
20107,7950119,4080130,89110142,29140153,58
21108,1851119,7881131,27111142,66141153,95
22108,5752120,1682131,66112143,04142154,32
23108,9653120,5583132,04113143,42143154,70
24109,3554120,9384132,42114143,80144155,07
25109,7355121,3285132,80115144,17145155,45
26110,1256121,7086133,18116144,55146155,82
27110,5157122,0987133,56117144,93147156,19
28110,9058122,4788133,94118145,31148156,57
29111,2859122,8689134,32119145,68149156,94